Comparison and evaluation of surface deformation of Hyflex controlled memory and Hyflex electric discharge machining nickel titanium rotary files and cyclic fatigue resistance after instrumentation and heat sterilization – An in vitro study


Department of Conservative Dentistry and Endodontics, Dayananda Sagar College of Dental Sciences, Bengaluru, Karnataka, India

Click here for correspondence address and email

Date of Submission29-Jan-2020
Date of Decision25-Jun-2020
Date of Acceptance03-Jul-2020
Date of Web Publication04-Aug-2020
 

   Abstract 

Aims: The aim of this study is to evaluate and compare the surface deformation and cyclic fatigue resistance of Hyflex controlled memory (CM) and Hyflex electric discharge machining (EDM) files following instrumentation and heat sterilization.
Subjects and Methods: Thirty Hyflex CM and thirty Hyflex EDM files were selected and profilometric images taken. Sixty extracted molars were decoronated, cleaning and shaping done with files, and subjected to autoclaving. Profilometric images were taken after instrumentation and sterilization. Cyclic fatigue testing was done at 30° and 60° in a custom-made jig.
Statistical Analysis Used: The Mann–Whitney test was used to compare profilometer and cyclic fatigue values between Hyflex CM and Hyflex EDM groups, whereas Friedman's and Wilcoxon signed-rank test were used for intragroup comparisons.
Results: Intact Hyflex EDM files showed higher surface roughness values compared to Hyflex CM. Hyflex CM showed an increase in surface roughness after instrumentation and sterilization that was statistically significant while Hyflex EDM showed no statistical significance both after instrumentation and sterilization. Hyflex EDM showed a significantly higher mean number of cycles to failure than Hyflex CM at both degrees.
Conclusions: Hyflex EDM files are able to maintain their surface topography and also have cyclic fatigue resistance better than Hyflex CM after both instrumentation and sterilization.

Keywords: Cyclic fatigue; Hyflex controlled memory; Hyflex electric discharge machining; optical profilometry; sterilization

How to cite this article:
Das P, Swapna D V, Nadig RR. Comparison and evaluation of surface deformation of Hyflex controlled memory and Hyflex electric discharge machining nickel titanium rotary files and cyclic fatigue resistance after instrumentation and heat sterilization – An in vitro study. J Conserv Dent 2019;22:464-9

How to cite this URL:
Das P, Swapna D V, Nadig RR. Comparison and evaluation of surface deformation of Hyflex controlled memory and Hyflex electric discharge machining nickel titanium rotary files and cyclic fatigue resistance after instrumentation and heat sterilization – An in vitro study. J Conserv Dent [serial online] 2019 [cited 2022 May 19];22:464-9. Available from: https://www.jcd.org.in/text.asp?2019/22/5/464/291328

   Introduction Top


The introduction of nickel titanium (NiTi) rotary instruments (1988) for endodontic procedures has made root canal therapy simpler and faster.[1] Despite all the advantages, the possibility of unexpected fracture of NiTi rotary files within canals because of cyclic/flexural and torsional fatigue distresses clinicians.[2] To prevent this procedural complication, manufacturers have come up with various methods to improve the mechanical properties and development of newer alloys is one among them.[3]

A thermally treated controlled memory (CM) alloy has been introduced recently in which the austenite/martensite transition temperature has been shifted so that a stable martensite microstructure is obtained at body temperature (B19´ martensite + R-Phase + Austenite).[4],[5] This removes the super elasticity property of conventional NiTi instruments.[6] Thus, the induced fracture due to torsion or flexural fatigue is prevented.

Hyflex CM files were introduced (Coltene Whaledent Alstatten, Switzerland) based on a special thermomechanical procedure. These files claim improved flexibility, resistance to fatigue compared to conventional rotary instruments with added advantage of reversion achieved by heat sterilization. Clinically, these CM rotary instruments have shown outstanding cyclic fatigue resistance.[4],[7]

Hyflex EDM (Coltene Whaledent, Alstatten, Switzerland) instruments are manufactured through electrodischarge machining process. This causes a superficially isotropic surface on the material, which is said to increase the cyclic fatigue and torsional stress resistance of the instrument.[8] Due to the unique control memory mechanism of this instrument, it is easier to negotiate complex canal curvatures causing less iatrogenic complications and also exhibits reversion on heat sterilization.[4],[7] The manufacturer also claims that the cyclic fatigue resistance of Hyflex CM is 300% compared to Hyflex EDM which is around 700% thus reducing the risk of fatigue failure.[9]

Due to the various manufacturing process, the incorporation of surface microcracks and cavities in the grooves and cutting edges of the files are unavoidable which, in turn, compromises the cyclic and torsional fatigue resistance of the files.[10] In addition, the chances of failure increase exponentially when the files are rotated inside a severely curved canal.[11] In a study, it was reported that 93% of instrument failures were due to this type of induced fatigue.[12]

The surface characteristics of NiTi instruments subjected to cyclic fatigue were determined mainly using scanning electron microscopy (SEM) or atomic force microscopy (AFM).[13],[14] Recently, Ferreira et al. suggested the use of optical profilometry in analyzing the surface properties of endodontic files because this method makes it possible to analyze wider surfaces with reproducible measurements and superior qualitative and quantitative information when compared with AFM and SEM.[15]

On reviewing the literature, studies examining the preuse, postuse, and postheat sterilization surface properties and topographies of Hyflex CM and Hyflex EDM files through three-dimensional (3D) optical profilometry and cyclic fatigue resistance of these instruments after instrumentation and subsequent heat sterilization were not available. Hence, the present study was undertaken to evaluate them.


   Subjects and Methods Top


Sixty intact extracted permanent molars with root curvature of 30°–60° (Schneider method) and with complete root formation were collected according to CDC guidelines. Thirty new Hyflex CM and thirty new Hyflex EDM instruments were tested for the biomechanical preparation of curved root canals. The instruments selected were 21 mm and 0.25 mm tip size (Hyflex CM 0.06 taper, and Hyflex EDM ~ taper). Before the experiment, the files were inspected using a stereomicroscope and any file with defect was discarded.

The instruments were grouped as follows (n = 30):

  • Group A: Hyflex CM
  • Group B: Hyflex EDM.


Teeth were decoronated 2 mm above the cemento-enamel junction (CEJ) to provide straight line access and stable reference plane. Working length was calculated 1 mm short of the apical foramen.

Surface roughness and deformation

The intact files from both the groups were evaluated under optical profilometer before instrumentation, and the images were recorded 3 mm from the tip of each instrument. Coronal preflaring was done with 25/0.08 file in the Hyflex CM group and with 25/0.12 file for the Hyflex EDM group. Cleaning and shaping were conducted with all the experimental files in compliance with the manufacturer's recommendation. The canals were irrigated with 3% sodium hypochlorite and 17% ethylene diamine tetra acetic acid (EDTA).

After instrumentation, the experimental files were observed under the optical profilometer for surface roughness, deformations of the cutting blades and flutes, and the values were recorded for both. Instruments of both the experimental groups were cleaned and subjected to autoclave sterilization at 121°C at 15 psi for 15 min and observed again under the optical profilometer to assess the change in surface roughness, reversion of the cutting blades and flutes, and the values were recorded.

Cyclic fatigue testing

Group A and Group B files were further subdivided into two groups of 15 each as A1 and B1 (30°) and A2 and B2 (60°). Files were tested inside a custom-made stainless steel artificial canal having a curvature of 30° or 60° with a 5 mm radius of curvature in the laboratory. The diameter (1.5 mm) of the simulated canals was greater than that of the instruments tested, thus allowing free rotation.

All the instruments were rotated according to the manufacturers recommended settings (speed: 500 rpm Torque: 2.5 Ncm) using a 16:1 gear reduction hand piece (X-smart, DENTSPLY) powered by a torque controlled electric motor until failure occurred. To reduce friction on the files and prevent overheating while rotating inside the curved canals, lubricating oil (WD-40, Milton Keynes, UK) was used.

A stop clock was used to measure the time duration from initiation of the cycles to failure. The time to fracture in seconds was multiplied by rotations per minute to obtain the number of cycles to failure (NCF) for each instrument. The length of each fractured instrument was measured using a digital caliper.


   Results Top


The statistical analyses were performed using the SPSS software version 22.0, 2013 (IBM Corp., Armonk, NY, USA). The Mann–Whitney test was used to compare profilometer and cyclic fatigue values between Hyflex CM and Hyflex EDM groups. Friedman's Test and Wilcoxon Signed-rank tests were used to compare the values between different time intervals and different curvatures. The level of significance was set at P < 0.05.

The parameters evaluated are as follows:

  • Sa = Mean of the surface roughness
  • Sq. = Square root of the mean of the surface roughness
  • Sz = Average height between peaks and valleys in the optical measurement field.


The mean and standard deviations of the Sa, Sq, and Sz values are shown in [Table 1]. The Sa, Sq, and Sz values of the HyFlex EDM group were significantly higher than those of the HyFlex CM group before use, after use and after sterilization (P < 0.05). In both groups, the tested amplitude parameters (Sa, Sq, and Sz) increased after root canal preparation. These parameters were statistically significantly increased in the HyFlex CM group after use and sterilization (P < 0.05), but there was no significant difference between the after use and post sterilization groups (P > 0.05). In contrast, the increase in the HyFlex EDM group was not statistically significant both after use and sterilization (P > 0.05).
Table 1: Mean and Standard Deviations of Sa, Sq, and Sz parameters before use, after use, and after sterilization of HyFlex CM and Hyflex EDM NiTi Instruments (in μm)

Click here to view


[Figure 1] and [Figure 2] show the optical profilometric images of Hyflex CM and Hyflex EDM files before use, after use, and after sterilization.
Figure 1: Profilometric image of Hyflex controlled memory files: (a) Before use; (b) After use; (c) After sterilization

Click here to view
Figure 2: Profilometric image of Hyflex electric discharge machining files: (a) Before use; (b) After use; (c) After sterilization

Click here to view


The mean NCF and the length of fractured instruments in the cyclic fatigue testing are in [Table 2]. The mean NCF was higher and the mean length of fractured segment was smaller for Hyflex EDM than Hyflex CM and this was statistically significant (P < 0.05).
Table 2: Mean number of cycles to failure and mean length of fractured fragment (in mm) between different curvatures in Hyflex CM and Hyflex EDM groups

Click here to view



   Discussion Top


Hyflex CM and Hyflex EDM files are manufactured from the same CM wire but using different processes. Hyflex CM is manufactured using thermomechanical process during which there are the formation of surface irregularities such as pits, fissures, and metal folds increasing the surface roughness. This may act as a starting point for crack initiation which might progress during clinical usage causing fracture of the files due to cyclic fatigue.[16],[17] This can be a reason for the increase in postinstrumentation roughness (Sa) seen in this study.

The Sz values of Hyflex CM increased significantly after use and when poststerilization files were compared with intact file there was statistically significant difference, implying that Hyflex CM was not able to completely revert back to its original preuse configuration after heat treatment. Otsuka and Ren also reported that the transformation temperatures of used CM wire required to revert back to its original configuration increases with usage when compared to unused wire due to progressive accumulation of mechanical stress.[18]

According to the qualitative parameters, intact Hyflex EDM files were found to have a significantly rougher surface than intact Hyflex CM files owing to their manufacturing process (electrical discharge machining). However, they showed no significant increase in the parameters evaluated both after use and sterilization. EDM technology perhaps helped in the better preservation of its surface characteristics and surface roughness values.[19] It may also be attributed to the unique white layer (SEM) formed on its superficial surface during manufacturing, which has higher hardness and better properties compared to the bulk material.[8]

In addition to the EDM manufacturing process, another reason for the better performance of Hyflex EDM could be the difference in usage instructions recommended by the manufacturer. The larger coronal preflaring with Hyflex EDM 25/0.12 compared with Hyflex CM may have resulted in better preservation of Hyflex EDM 25/0.08 files' surface properties.[19]

In parallel with the present study, Uslu et al. also compared the alterations in the surface topographies of Hyflex CM and Hyflex EDM files before and after root canal preparation using 3D optical profilometry, and the results were in accordance with the results obtained in our study.[19]

Lopes et al. reported that the grooves on the cutting edges of files have negative effect on the cyclic fatigue resistance and might be responsible for their lower cyclic fatigue resistance.[20] Hence, this study also assessed the cyclic fatigue fracture of these instruments after instrumentation and heat sterilization.

Pruett et al. showed that the instruments separate at the point of maximum flexure of the shaft, which corresponds to the midpoint of curvature.[21] In this study, Hyflex CM fractured at a range of 6.8–7.1 mm at 30° and 60° which was statistically significant implying that the point of flexure changes for the file when subjected to different curvatures. For Hyflex EDM, the fracture occurred at a range of 5.5–5.7 at 30° and 60°, and this was not statistically significant showing that the point of flexure remains the same for the file irrespective of the angle of curvature. This could be because the taper of Hyflex EDM decreases from 8% to 4%, 4 mm from the tip of the file where it is subjected to maximum flexural stress.[12]

The mean NCF was higher for Hyflex EDM than Hyflex CM at both angles, and this difference was statistically significant. This may be attributed to the better preservation of surface topography and better fatigue life of EDM files.

Under the conditions of this study, fracture and overall failure were more frequent in severely curved canals (60°) than in moderately curved canals (30°). This can be explained by the overloading induced on the instrument by an abrupt change in canal curvature which restrains the rotating instrument, giving rise to multidirectional loading (tension, bending and torsion) that lead to fracture of the instruments.[22] An increase in angle of curvature from 30° to 60° led to significant increase in fracture incidence for both file systems which was statistically significant.


   Conclusions Top


Within the limitations of the present study, Hyflex CM files showed a significantly higher level of higher surface roughness and deformation compared with the Hyflex EDM files after the preparation of severely curved canals and sterilization. EDM technology provided better preservation of NiTi file surface after clinical usage in severely curved canals when compared with conventional grinding method. Hyflex EDM files have a better resistance to cyclic fatigue at both 300 and 600 curvatures than Hyflex CM.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Peters OA. Current challenges and concepts in the preparation of root canal systems: A review. J Endod 2004;30:559-67.  Back to cited text no. 1
    
2.
Kim HC, Yum J, Hur B, Cheung GS. Cyclic fatigue and fracture characteristics of ground and twisted nickel-titanium rotary files. J Endod 2010;36:147-52.  Back to cited text no. 2
    
3.
Mavani P, Pujar M, Uppin V, Vagarali H, Patil C, Yalagi V. Comparative evaluation of root cracks by different rotary and reciprocating endodontic file systems. IOSR J Dent Med Sci 2015;14:18-22.  Back to cited text no. 3
    
4.
Santos Lde A, Bahia MG, de Las Casas EB, Buono VT. Comparison of the mechanical behavior between controlled memory and superelastic nickel-titanium files via finite element analysis. J Endod 2013;39:1444-7.  Back to cited text no. 4
    
5.
Zhou HM, Shen Y, Zheng W, Li L, Zheng YF, Haapasalo M. Mechanical properties of controlled memory and superelastic nickel-titanium wires used in the manufacture of rotary endodontic instruments. J Endod 2012;38:1535-40.  Back to cited text no. 5
    
6.
Young JM, Van Vliet KJ. Predictingin vivo failure of pseudoelastic NiTi devices under low cycle, high amplitude fatigue. J Biomed Mater Res B Appl Biomater 2005;72:17-26.  Back to cited text no. 6
    
7.
Pirani C, Iacono F, Generali L, Sassatelli P, Nucci C, Lusvarghi L, et al. HyFlex EDM: Superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int Endod J 2016;49:483-93.  Back to cited text no. 7
    
8.
Niladri M and Swati M. The New View towards EDM Files. Journal of Dental Sciences 2017;2. DOI:10.23880/oajds-16000144.  Back to cited text no. 8
    
9.
Hyflex ED brochure. Coltene/Whaledent GmbH + Co. KG, Langenau: Germany; 2015.  Back to cited text no. 9
    
10.
Eggert C, Peters O, Barbakow F. Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopy. J Endod 1999;25:494-7.  Back to cited text no. 10
    
11.
Alapati SB, Brantley WA, Svec TA, Powers JM, Mitchell JC. Scanning electron microscope observations of new and used nickel-titanium rotary files. J Endod 2003;29:667-9.  Back to cited text no. 11
    
12.
Cheung GS, Peng B, Bian Z, Shen Y, Darvell BW. Defects in ProTaper S1 instruments after clinical use: Fractographic examination. Int Endod J 2005;38:802-9.  Back to cited text no. 12
    
13.
Valois CR, Silva LP, Azevedo RB. Atomic force microscopy study of stainless-steel and nickel-titanium files. J Endod 2005;31:882-5.  Back to cited text no. 13
    
14.
Valois CR, Silva LP, Azevedo RB. Multiple autoclave cycles affect the surface of rotary nickel-titanium files: An atomic force microscopy study. J Endod 2008;34:859-62.  Back to cited text no. 14
    
15.
Ferreira F, Barbosa I, Scelza P, Russano D, Neff J, Montagnana M, et al. A new method for the assessment of the surface topography of NiTi rotary instruments. Int Endod J 2017;50:902-9.  Back to cited text no. 15
    
16.
Pook LP. Metal fatigue-what it is, why it matters. Netherlands: Springer; 2007. p. 161-5.  Back to cited text no. 16
    
17.
Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35:1469-76.  Back to cited text no. 17
    
18.
Ostuka K, Ren X. Physical metallurgy of Ti-Ni-based shape memory alloys. Progress Materials Sci 2005;50:511-678.  Back to cited text no. 18
    
19.
Uslu G, Özyürek T, Yılmaz K. Comparison of alterations in the surface topographies of hyflex CM and hyflex EDM nickel-titanium files after root canal preparation: A three-dimensional optical profilometry study. J Endod 2018;44:115-9.  Back to cited text no. 19
    
20.
Lopes HP, Elias CN, Vieira VT, Moreira EJ, Marques RV, de Oliveira JC, et al. Effects of electropolishing surface treatment on the cyclic fatigue resistance of BioRace nickel-titanium rotary instruments. J Endod 2010;36:1653-7.  Back to cited text no. 20
    
21.
Pruett JP, Clement DJ, David L, Carnes Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23:77-85.  Back to cited text no. 21
    
22.
Spanaki-Voreadi AP, Kerezoudis NP, Zinelis S. Failure mechanism of ProTaper NiTi rotary instruments during clinical use: Fractographic analysis. Int Endod J 2006;39:171-8.  Back to cited text no. 22
    

Top
Correspondence Address:
Dr. Paramarshi Das
Department of Conservative Dentistry and Endodontics, Dayananda Sagar College of Dental Sciences, 1st Stage, Shavige Malleshwara Hills, Kumaraswamy Layout, Bengaluru - 560 078, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JCD.JCD_39_20

Rights and Permissions


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


    Abstract
   Introduction
   Subjects and Methods
   Results
   Discussion
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1582    
    Printed49    
    Emailed0    
    PDF Downloaded81    
    Comments [Add]    

Recommend this journal