ORIGINAL ARTICLE
Year : 2019  |  Volume : 22  |  Issue : 5  |  Page : 449-453

Influence of different motions on the cyclic fatigue resistance of Reciproc and Reciproc Blue endodontic instruments


1 Private Practice, Samarate, Italy
2 Clinical Department of Medical, Surgical and Health Sciences, University of Trieste, Trieste, Italy
3 Private Practice, Rome, Italy
4 Department of Surgery, Medicine, Dentistry and Morphological Sciences with Transplant Surgery, Oncology and Regenerative Medicine Relevance, University of Modena and Reggio Emilia, Modena, Italy

Correspondence Address:
Prof. Daniele Angerame
Clinical Department of Medical, Surgical and Health Sciences, University of Trieste, Piazza Ospedale 1, 34129 Trieste
Italy
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JCD.JCD_430_19

Rights and Permissions

Aim: The aim of this study was to evaluate the cyclic fatigue resistance of Reciproc (RCP) and RPC Blue (RCPB) instruments used in continuous rotation, “RECIPROC” mode, and “WAVEONE” mode. Materials and Methods: Sixty RCP and 60 RCPB R25 files were used. For each file type, three groups (n = 20) were defined depending on the used kinematics: continuous rotation, “RECIPROC” mode, and “WAVEONE” mode. A stainless-steel artificial canal with 60° angle and 5-mm radius of curvature was milled reproducing the size and taper of the used files. The test device was electrically heated to 35°C to simulate the clinical environment. All files were reciprocated or rotated until fracture. The time to failure and the length of the fractured fragments were measured. A fractographic examination was performed by scanning the electron microscopy to confirm the cause of fracture. Collected data underwent a two-way analysis of variance (α = 0.05). Results: RCPB files exhibited better cyclic fatigue resistance than RCP. The “RECIPROC” motion yielded greater cyclic fatigue resistance than the “WAVEONE” movement; the least resistance was observed in the continuous rotation groups. No significant differences were found among groups in terms of length of the fractured fragment. The fractographic analysis confirmed that all scanned samples separated due to cyclic fatigue. Conclusions: Within the limitation of the present study, the “RECIPROC” mode increased the cyclic fatigue resistance of the tested instruments compared to “WAVEONE” mode and continuous rotation. To prevent RCP and RCPB file separation, motion kinematics other than the native “RECIPROC” movement should be discouraged in the clinical setting.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1272    
    Printed47    
    Emailed0    
    PDF Downloaded70    
    Comments [Add]    
    Cited by others 2    

Recommend this journal